Hi, ada yang dapat kami bantu?

Radwag PUE 5.19IR Terminal

OVERVIEW

The terminals are intended to designing scales with maximum resolution of 6000 e. PUE 5 series, similarly to the basic model, are enclosed in a stainless steel housing which allows for reaching the highest IP rating, i.e. IP68. The terminals come standard with interfaces: 2 × RS232, 1 × RS485, 1 × Ethernet, 4 × USB 2.0. PUE 5 series can be extended by additional 3 weighing modules for simultaneous operation of maximum 4 weighing platforms.

Optionally available PROFIBUS DP interface enables terminals cooperation with industrial ProfiBus network.

Intended Use:

The terminal PUE 5 series is a multifunctional weighing device. It combines the best features of a PC computer and a scale. The PUE 5 operates on basis of OS Windows, creating the best option for configuring the device to individual customer requirements through installing corresponding application.

If applied with dosing or formula making systems, the PUE 5 allows for operation with multiple weighing platforms, and simultaneous dosing with full process control. A dosing or formula making system that is configured on basis of the terminal PUE 5 series does not require an additional PC computer. Database editing is possible from the level of PUE 5 touch screen display or an optional PC keyboard.

Area of Use:

Main applications for terminal PUE 5 series:

  • Designing industrial load cell scales;
  • Multiple weighing applications supervised by a computer system through Ethernet module;
  • Possibility of connecting large quantity of additional weighing platforms to industrial terminals.
Terminal types:
  • PUE 5.19IR – infrared touch screen technology. Protection class IP68. Very high brightness and readability level. High resistance to moisture and dust. High mechanical resistance, possibility of operation using gloves (made of latex, material or even steel chain).

Butuh Bantuan?

Informasi Produk

FUNCTIONS

DOSING

Weighing process for which reference mass has been determined together with tolerance for its determination. Dosing tolerance is given in [%] and it is calculated in relation to the reference value thus being a permissible deviation of this process. This solution is used for weighing powders, liquids and loose materials. Dosing function performance is often supported with bargraph – load indicator. For industrial scales it is possible to use a control systems of dosing process.

LABELLING

Marking products with the right symbols contributes to quick and reliable identification. The balance/scale usually transfers product mass or product quantity to the label printer. The transferred value is printed with the following information: bar code, pictogram, content, address etc. In laboratory the labelling is performed in a semi-automatic mode, for industry in an automatic one (dynamic scales).

PLUS/MINUS CONTROL

Control of sample’s mass carried out with reference to specified MIN and MAX thresholds. The weight is considered to be correct if it is comprised within threshold values. A graphic visualisation of weighing result is displayed in a form of a bar graph.

PARTS COUNTING

Function using mass measurement for determination of  measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.

FORMULATION

Function supporting the mixture making process, wherein the mixture contains various components. Formulation function usually uses the balance/scale database of components. Formulation serves for monitored checkweighing of every single component with a given tolerance. It is enriched with a set of individual settings.

STATISTICS

Statistiscs function registers and analyses performed measurements. This supplies the user withthe following information: Max and Min standard deviation, average value, variance, range et.

CHECKWEIGHING

Checkweighing function is used for checking whether the measured sample mass is within the predefined threshold values, Low [LO] and Hgh [HI]. The thresholds are given in [g] and [kg] units. Current state of a sample being measured is signalled by means of pictorgams located on a display for laboratory balances, for industrial scales Stackligt System is used. This visual +/- inspection is in operation during segregation, control or packing process of products for which mass has bees determined with a specifaied tolerance, eg. 12860 g 961

DATASHEET       

Maximum quantity of verification units 6000 e
OIML Class III
Display 19” graphic colour touchscreen
Protection class IP 68
USB Interface 4×USB 2.0 (2×USB-A, 2×M12 4P)
Ethernet 2×10/100/1000 Mbit (2×RJ45 hermetic)
Profibus DP Interface operation mode: slave
Serial Interface 2×RS232, RS485 (2×M12 8P)
Inputs/Outputs (option) 4IN / 4OUT (2×ports M12 8P or 2×cables fed through the cable gland)
Additional weighing platform module additional max. 3 pcs (metrological parameters as in the main weighing platform)
External buttons PRINT, TARA, START (option) terminal with 4IN/4OUT (M12 8P connector) is required
Power consumption max. 50 W
Maximum quantity of divisions from converter 838 860 ×10
Maximum increase of signal 19.5 mV
Minimum volatge per verification unit 0.4 µV
Maximum voltage per verification unit 3.25 µV
Minimum load cell impedance 80 Ω
Maximum load cell impedance 1200 Ω
Load cell excitation voltage 5V DC
Connection of load cells 4 or 6 wires + shield
Operating temperature 0 ÷ +40°C
Relative humidity 10% ÷ 85% RH no condensation
Touch panel infrared
Housing stainless steel
Operating system Microsoft Windows 10 IoT
Processor Intel® Pentium® N3710 on-board SoC (2.56GHz, quad-core, 2MB cache)
Chipset Intel® HD Graphics Gen 7
Memory 4GB 1066/1333 MHz DDR3L max 8 GB
Data memory SSD 128 GB
Packaging dimensions 590×430×240 mm
Net weight 14 kg
Gross weight 15 kg

COMPATIBLE WITH

  • F1 Mass Standard-knob weights without adjusment chamber, set
  • Stainless steel HRP. H Platforms
  • HRP Platforms High Capacity
  • Waterproof  stainless steel platforms
  • Mild steel powder coated weighing platforms
  • Platforms in plastic casing
  • Waterproof  stainless steel platforms
  • Stainless steel ramp platforms with LD lifting system
  • Ramp stainless steel platforms
  • Stainless steel platforms, pit version
  • Mild steel powder coated weighing platforms
  • E2R System
  • R-LAB
  • RAD-KEY
  • labVIEW Driver
  • Receipt Printer
  • Label Printers
  • Moxa Uport 1110 Converter
  • RS 232 cables (scale – ZEBRA printer)
  • USB Adapter
  • USB cables (terminal – terminal)
  • IN/OUT Cables
  • RS 232 cables (scale- EPSON printer)
  • USB cables (scale – Ethernet)
  • USB cable (scale – printer)
  • RS 232, RS 485 cables
  • Keypad, external switches
  • Barcode Scanners
  • Displays