Radwag C315.300.C2.M Load Cell Platform Scale

OVERVIEW

Maximum capacity [Max]: 300kg

Readability [d]: 100g
Platform scales C315 series is designed for direct trade. It is used wherever fast and precise determination of net weight is required.

The C315 series features an indicator PUE C315 series comprising backlit LCD display.

C315 scales is manufactured using powder coated mild steel profiles, with stainless steel weighing platform.

Scales C315 series is available in dual range version on surcharge.


 

Butuh Bantuan?

Informasi Produk

FUNCTIONS

PLUS/MINUS CONTROL

Control of sample’s mass carried out with reference to specified MIN and MAX thresholds. The weight is considered to be correct if it is comprised within threshold values. A graphic visualisation of weighing result is displayed in a form of a bar graph.

PERCENT WEIGHING

Percent Weighing function is used for comparision of measured products with mass standard. Mass of a mass standard may be a numeric value taken from a database or it may be determined through a measurement process. Each measured product is compared to mass standard, mass of which is presumed as a model 100% ideal mass. For products weighing less than the mass standard, obtained results are lower than 100%, for products weighing more, the obtained results are greatly  exceeded.

TOTALIZING

Function adding new mass value to already obtained one. This allows to determine mass content of the mixture, being a composition of number of different consistence components. Totalizing is limited by maximum capacity parameter.

PARTS COUNTING

Function using mass measurement for determination of  measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.

INTERNAL BATTERY

Internal and rechargeable battery of SLA 6 V type (sealed lead acid) enables device operation without connecting the device to the mains (9h operation). The weighing instrument controls battery status and displays a respective pictogram on the screen.

PEAK HOLD

Function registering the greatest temporary indication occurring in course of the weighing process. It is most often used in medical scales in order to hold the measurement indication.

GLP PROCEDURES

Diagnostic function allowing to objectively document performed measurements. GLP procedures may be either presented in a short report form or extended one.

ANIMAL WEIGHING

Process of mass determination for a product which may unwillingly reposition within the weighing pan. Mass determination in such cases requires much longer period of time when compared to typical weighing process. It is the user who defines period of time needed for control of measured product mass. The user can thus optimize the function depending on the measured product characteristics.

ALIBI MEMORY

“ALIBI” memory enables saving and storing up to 100000 measurements.  Measurements are saved automatically upon each pressing of the key. There is no need for any additional operations or changing settings.
The following data related with the measurements is saved: measurement date; measurement time; result (mass); tare value; operator carrying out the measurement (logged in); product name.
The software saves the measurement in a so called loop, i.e. when the measurement no. 100001 is saved, the measurement 1 gets automatically deleted from the memory.
Measurements saved to balance memory cannot be deleted.
You can preview and print out the data saved to the ALIBI database.
DATASHEET 
Maximum capacity [Max]300 kg
Minimum load2 kg
Readability [d]100 g
Tare range-300 kg
Verification scale interval100 g
Max readability for non-verified scale10 g
OIML ClassIII
DisplayLCD (backlit)
Protection classIP 65 construction, IP 43 indicator
RS232 Interface1× standard, 1× option
Power supply100 ÷ 240 V AC 50 / 60 Hz and battery
Power consumption max.5 W
Operation time on batteriesmax 8h
Operating temperature-10 ÷ +40 °C
Storage temperature-10 ÷ +50 °C
Relative humidity10% ÷ 85% RH no condensation
IndicatorC315
Constructionmild steel St3S
Weighing pan materialstainless steel AISI304
Weighing pan dimensions400×500 mm
Packaging dimensions720×620×210 mm
Net weight12.8 kg
Gross weight14.8 kg

ACCESSORIES

  • Granite Antivibration tables
  • Scales Editor 2.0
  • RW Tool
  • R Panel
  • R-Lab
  • RAD – KEY
  • LabVIEW Driver
  • Receipt Printer
  • Moxa Uport 1110 Converter
  • RS232 – Ethernet Converter
  • RS232 1345 Converter
  • RS 232 cables (scale – EPSON printer)
  • RS 232, RS 485 cables
  • Cigarrete lighter receptacle power supply cables
  • Displays

Our customer support team is here to answer your questions. Ask us anything!
//
Jakarta, Banten, Sumatra & Sulawesi
ABDUL SYUKUR
Available
//
Jawa Barat
BAYU SOFYAN
Available
//
Jawa Timur & Jawa Tengah
MARYOTO
Available
NOTICE
We use cookies to make your experience of our websites better.