FUNCTIONS
LABELLING
Marking products with the right symbols contributes to quick and reliable identification. The balance/scale usually transfers product mass or product quantity to the label printer. The transferred value is printed with the following information: bar code, pictogram, content, address etc. In laboratory the labelling is performed in a semi-automatic mode, for industry in an automatic one (dynamic scales).
PLUS/MINUS CONTROL
Control of sample’s mass carried out with reference to specified MIN and MAX thresholds. The weight is considered to be correct if it is comprised within threshold values. A graphic visualisation of weighing result is displayed in a form of a bar graph.
PERCENT WEIGHING
Percent Weighing function is used for comparision of measured products with mass standard. Mass of a mass standard may be a numeric value taken from a database or it may be determined through a measurement process. Each measured product is compared to mass standard, mass of which is presumed as a model 100% ideal mass. For products weighing less than the mass standard, obtained results are lower than 100%, for products weighing more, the obtained results are greatly exceeded.
PARTS COUNTING
Function using mass measurement for determination of measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.
STATISTICS
Statistiscs function registers and analyses performed measurements. This supplies the user withthe following information: Max and Min standard deviation, average value, variance, range et.
IR SENSORS
Programmable function supporting the weighing process through control of the following options: sliding weighing chamber doors, printout, zeroing, tarring etc. Especially appreciated wherever preventing the balance from soiling is important.
UNDER-PAN WEIGHING
The solution enables weighing with use of a stand, which is advisable for loads of non-standard shapes and dimensions and those that generate magnetic field.
GLP PROCEDURES
Diagnostic function allowing to objectively document performed measurements. GLP procedures may be either presented in a short report form or extended one.
REPLACEABLE UNIT
Replaceable units.
ALIBI MEMORY
“ALIBI” memory enables saving and storing up to 100000 measurements. Measurements are saved automatically upon each pressing of the key. There is no need for any additional operations or changing settings.
The following data related with the measurements is saved: measurement date; measurement time; result (mass); tare value; operator carrying out the measurement (logged in); product name.
The software saves the measurement in a so called loop, i.e. when the measurement no. 100001 is saved, the measurement 1 gets automatically deleted from the memory.
Measurements saved to balance memory cannot be deleted.
You can preview and print out the data saved to the ALIBI database.
The following data related with the measurements is saved: measurement date; measurement time; result (mass); tare value; operator carrying out the measurement (logged in); product name.
The software saves the measurement in a so called loop, i.e. when the measurement no. 100001 is saved, the measurement 1 gets automatically deleted from the memory.
Measurements saved to balance memory cannot be deleted.
You can preview and print out the data saved to the ALIBI database.
DATASHEET
Maximum capacity [Max] | 60 kg |
Minimum load | 400 g |
Readability [d] | 20 g |
Tare range | -60 kg |
OIML Class | III |
Display | 5” graphic colour |
Protection class | IP 65 construction, IP 43 indicator |
Communication interface | 2×RS232, USB-A, USB-B, Ethernet, 4 IN / 4 OUT (digital), Wi-Fi® |
Power supply | 100 ÷ 240 V AC 50 / 60 Hz and battery* |
Power consumption max. | 10 W |
Operation time on batteries | operation time up to 5 h |
Operating temperature | -10 ÷ +40 °C |
Construction | mild steel St3S |
Weighing pan material | stainless steel AISI304 |
Weighing pan dimensions | 500×400 mm |
Packaging dimensions | 720×620×210 mm |
Net weight | 12.8 kg |
Gross weight | 14.8 kg |
ACCESSORIES
- E2R System
- Label Editor R02
- R.Barcode
- R-Lab
- RAD-KEY
- RADWAG Connect
- RADWAG Development Studio
- LabVIEW Driver
- Receipt Printer
- Cigarrete lighter receptacle power supply cables
- RS 232, RS 485 cables
- RS 232 cables (scale – EPSON printer)
- Keypad, external switches
- Barcode Scanners
- Stands, wall mounting kits and mounting brackets
- Displays