Databases:
Information system of R series moisture analyzers is based on 6 databases, allowing many operators to operate product database comprising many samples. Collected measurements may be subjected to subsequent analysis.
Collected data is registered in 6 databases:
- Users (up to 10 users)
- Products (up to 1000 products)
- Weighings (up to 1000 weighings)
- Tares (up to 100 tares)
- Programs (up to 100 drying programs)
- Drying process reports (up to 1000 reports)
Exchange of information within the system is performed by means of USB port on bi-directional basis. It is possible to import and export databases using memory sticks.
Direct access to information
It is possible to access functions and databases directly, using keyboard soft keys.
Database – access to databases
Function – access to basic functions
F1 and F2 – programmable function keys and menu navigation keys
Reports – access to database of drying process reports
Profile – access to parameters settings
Sample – access to products database
FUNCTIONS
DRYING MODES
SAMPLES DRYING
Drying process enables to determine moisture content of a given sample by evaporating its free water and other components. The sample’s moisture content [%M] is determined by means of precise weighing carried out before and during the drying process.

MOISTURE CONTENT ANALYSIS
DRY MASS DETERMINATION
DATASHEET
| Maximum capacity [Max] | 50 g |
| Readability [d] | 0.1 mg |
| Tare range | -50 g |
| Adjustment | external |
| Display | LCD (backlit) |
| Communication interface | RS232, USB-A, USB-B, Wi-Fi® |
| Power supply | 100 ÷ 240 V AC 50 / 60 Hz |
| Power consumption | 4 W |
| Operating temperature | +10 ÷ +40°C |
| Maximum sample weight | 50 g |
| Heating module | halogen |
| Heating module power | 450 W |
| Moisture content repeatability | +/-0.05% (sample ~ 2g), +/-0.01% (sample ~ 10g) |
| Moisture content readability | 0.0001% |
| Drying temperature range | max 250°C |
| Drying method | 4 drying profiles (standard, fast, step, mild) |
| Finish mode | 4 modes (automatic, manual, time-defined, user-defined) |
| Maximum sample height | 20 mm |
| Weighing pan dimensions | ø90, h= 8 mm |
| Packaging dimensions | 470×380×336 mm |
| Net weight | 4.8 kg |
| Gross weight | 6.3 kg |
| Additional functions | sample traceability |
ACCESSORIES
- Antivibration tables for laboratory balances
- ALIBI Reader
- E2R System
- R Panel
- RAD-KEY
- RADWAG Development Studio
- labVIEW Driver
- Moisture analyzer accessories
- Receipt Printer
- RS 232, RS 485 Cables
- RS 232 cables (scale-EPSON Printer)
- Protective Cover for Balances
- Barcode Scanners
DRYING PARAMETERS
Drying Parameters for MAC and MAX
Drying Parameters for MA 3Y, MA X2 and MA R
-
Chemical industry
- Fruit and vegetable
- Bakery industry
- Agricultural industry
- Food industry
-
Plastics
- Dairy products
- Other
Drying Parameters for PMV
- Dairy products
CAUTION
The analysis was carried out at customer request on selected models of moisture analyzers. RADWAG guarantees identical result of tests carried out on all versions of moisture analyzers.