Multifunctional Software
Indicator software allows carrying out processes such as weighing, parts counting, dosing, labelling, formulations and percent weighing. Indicator’s system of information is based on databases (such as: users, products, weighings, packaging, formulations, customers) and ALIBI memory which guarantees stored data safety.
When creating the software a great emphasis was placed on its functionality and ergonomics. This resulted in increased comfort of indicator operation.
Battery as an Optional Power Source
Use of the optional battery enables the PUE HX7 indicator to operate even when there is either no or unstable power supply. This improves the security of acquired data and improves comfort of operation.
Communication Interface
The indicator is equipped with RS 322, RS 485, USB, Ethernet, digital inputs/outputs and analog output. This enables cooperation with external devices: barcode scanners, printers, external displays, control buttons, light signalling towers, other controlling/signalling devices, systems for automatic process control and superior IT systems. It is possible due to the implemented and complex character-based communication protocol. In the future, the interfaces will be extended with Profibus and Profinet. Additionally, Modbus RTU and TCP protocols will be implemented.
FUNCTIONS
DOSING
Weighing process for which reference mass has been determined together with tolerance for its determination. Dosing tolerance is given in [%] and it is calculated in relation to the reference value thus being a permissible deviation of this process. This solution is used for weighing powders, liquids and loose materials. Dosing function performance is often supported with bargraph – load indicator. For industrial scales it is possible to use a control systems of dosing process.
PERCENT WEIGHING
Percent Weighing function is used for comparision of measured products with mass standard. Mass of a mass standard may be a numeric value taken from a database or it may be determined through a measurement process. Each measured product is compared to mass standard, mass of which is presumed as a model 100% ideal mass. For products weighing less than the mass standard, obtained results are lower than 100%, for products weighing more, the obtained results are greatly exceeded.
PARTS COUNTING
Function using mass measurement for determination of measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.
FORMULATION
Function supporting the mixture making process, wherein the mixture contains various components. Formulation function usually uses the balance/scale database of components. Formulation serves for monitored checkweighing of every single component with a given tolerance. It is enriched with a set of individual settings.
REPLACEABLE UNIT
ALIBI MEMORY
The following data related with the measurements is saved: measurement date; measurement time; result (mass); tare value; operator carrying out the measurement (logged in); product name.
The software saves the measurement in a so called loop, i.e. when the measurement no. 100001 is saved, the measurement 1 gets automatically deleted from the memory.
Measurements saved to balance memory cannot be deleted.
You can preview and print out the data saved to the ALIBI database.
DATASHEET
Maximum quantity of verification units | 6000 * |
OIML Class | III |
Display | 7” graphic colour |
Protection class | IP 66 / IP 68 |
USB Interface | 1× standard, 1× option (M12 4P) |
ETHERNET Interface | 1× standard |
EtherNet/IP Interface | 1 × option (RJ45) |
Profinet Interface | 1 × option (RJ45) |
RS232 Interface | 1× standard, 2× option |
1× option | |
IN/OUT Interface | 4 IN 4 OUT standard, 12 IN 12 OUT option |
Analog output module (option) | 4-20mA, 0-10V |
Power consumption max. | 25 W |
Optional power supply | internal rechargeable battery |
Operation time on batteries | max 7h |
Maximum increase of signal | 19.5 mV |
Minimum volatge per verification unit | 0.4 µV |
Maximum voltage per verification unit | 3.25 µV |
Minimum load cell impedance | 50 Ω |
Maximum load cell impedance | 1200 Ω |
Load cell excitation voltage | 5V DC
|
Connection of load cells | 4 or 6 wires + shield |
Operating temperature | -10 ÷ +40 °C |
Storage temperature | -10 ÷ +50 °C |
Relative humidity | 10% ÷ 80% RH no condensation |
Keypad | membrane |
Material | stainless steel AISI304 |
Device dimensions | 340×231×120 mm |
Packaging dimensions | 370×310×220 mm |
Net weight | 4.7 kg |
Gross weight | 5.5 kg |
Max number of platforms | max 2 (1× standard + 1× option) |
COMPATIBLE WITH
- Waterproof stainless steel platforms
- Mild steel powder coated weighing platforms
- Platforms in plastic casing
- Stainless steel platforms, pit version
- Stainless steel ramp platfoms with LD lifting system
- Ramp stainless steel platforms
- Waterproof stainless steel platforms
- Mild steel powder coated weighing platforms
- ALIBI Reader
- E2R System
- Label Editor R02
- R.Barcode
- R-LAB
- RAD-KEY
- RADWAG Connect
- RADWAG Development Studio
- labVIEW Driver
- Label Printers
- Moxa Uport 1110 Converter
- RS 232 cables (scale – ZEBRA printer)
- USB Adapter
- IN/OUT Cables
- USB cable (scale – printer)
- RS 232, RS 485 cables
- RS 232 cables (scale- EPSON printer)
- Barcode Scanners