Hi, ada yang dapat kami bantu?

Radwag PUE C315 Indicator

OVERVIEW

Functions of an indicator 
PUE C315 series:

  • measuring units: [g], [kg], [N], [ct], [lb];
  • tarring in whole measuring range;
  • automatic tare, tare memory;
  • parts counting with equal part mass;
  • +/- control of mass with set standard;
  • percent setup with set standard;
  • averaging of weighing
    result, digital filter;
  • control of power supply from batteries;
  • scale timer switch off;
  • adjustable backlight on powering by batteries;
  • adjustable baud rate between 1200-38400 bit/s;
  • continuous data transmission through RS 232;
  • manual or automatic mode for RS 232;
  • weighing loads with switched off autozero function;
  • measurement of maximum force or maximum mass
    placed on weighing pan;
  • measurement of force influencing the weighing pan (in Newtons);
  • start mass control;
  • slot for connecting an additional LCD display;
  • totalizing mode.

The indicator PUE C315 series enables designing printouts according to individual needs.

Butuh Bantuan?

SKU: c69de1b3baed Kategori:

Informasi Produk

FUNCTIONS

PLUS/MINUS CONTROL

Control of sample’s mass carried out with reference to specified MIN and MAX thresholds. The weight is considered to be correct if it is comprised within threshold values. A graphic visualisation of weighing result is displayed in a form of a bar graph.

PERCENT WEIGHING

Percent Weighing function is used for comparision of measured products with mass standard. Mass of a mass standard may be a numeric value taken from a database or it may be determined through a measurement process. Each measured product is compared to mass standard, mass of which is presumed as a model 100% ideal mass. For products weighing less than the mass standard, obtained results are lower than 100%, for products weighing more, the obtained results are greatly  exceeded.

TOTALIZING

Function adding new mass value to already obtained one. This allows to determine mass content of the mixture, being a composition of number of different consistence components. Totalizing is limited by maximum capacity parameter.

PARTS COUNTING

Function using mass measurement for determination of  measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.

INTERNAL BATTERY

Internal and rechargeable battery of SLA 6 V type (sealed lead acid) enables device operation without connecting the device to the mains (9h operation). The weighing instrument controls battery status and displays a respective pictogram on the screen.

PEAK HOLD

Function registering the greatest temporary indication occurring in course of the weighing process. It is most often used in medical scales in order to hold the measurement indication.

GLP PROCEDURES

Diagnostic function allowing to objectively document performed measurements. GLP procedures may be either presented in a short report form or extended one.

ANIMAL WEIGHING

Process of mass determination for a product which may unwillingly reposition within the weighing pan. Mass determination in such cases requires much longer period of time when compared to typical weighing process. It is the user who defines period of time needed for control of measured product mass. The user can thus optimize the function depending on the measured product characteristics.

ALIBI MEMORY

“ALIBI” memory enables saving and storing up to 100000 measurements.  Measurements are saved automatically upon each pressing of the key. There is no need for any additional operations or changing settings.

The following data related with the measurements is saved: measurement date; measurement time; result (mass); tare value; operator carrying out the measurement (logged in); product name.

The software saves the measurement in a so called loop, i.e. when the measurement no. 100001 is saved, the measurement 1 gets automatically deleted from the memory.

Measurements saved to balance memory cannot be deleted.

You can preview and print out the data saved to the ALIBI database.

DATASHEET

Maximum quantity of verification units 6000 e
OIML Class III or IV
Display LCD (backlit)
Protection class IP 43
RS232 Interface 1× standard, 1× option
Wi-Fi® Interface 1× standard
Power supply 100 ÷ 240 V AC 50 / 60 Hz
Power supply detachable adapter, akumulatory NiMH 6×AA
Optional power supply 12V DC and battery
Operation time on batteries max 7h
Maximum quantity of divisions from converter 838 860 ×10
Maximum increase of signal 19.2 mV
Minimum volatge per verification unit 0.4 µV
Maximum voltage per verification unit 3.25 µV
Minimum load cell impedance 50 Ω
Maximum load cell impedance 1200 Ω
Load cell excitation voltage 5V DC
Connection of load cells 4 or 6 wires + shield
Operating temperature -0,25
 °C
Storage temperature -0,2
 °C
Relative humidity 10% ÷ 80% RH no condensation
Keypad microswitch
Number of buttons 5
Housing ABS plastic
Device dimensions 181×136×60
 mm
Packaging dimensions 220×190×90
 mm
Net weight 00.06
 kg
Gross weight 01.02
 kg
Multiple range 1 or 2 ranges
Additional LCD display WD-4/1 (option)

COMPATIBLE WITH

  • Waterproof stainless steel platforms
  • Mild steel powder coated weighing platforms
  • Mild steel powder coated weighing platforms
  • Scales Editor 2.0
  • RW Tool
  • R Panel
  • R – LAB
  • RAD-KEY
  • RADWAG Connect
  • RADWAG Development Studio
  • labVIEW Driver
  • Receipt Printer
  • Label Printers
  • RS 232, RS 485 Cables
  • RS 232 cables (scale-EPSON Printer)
  • Moxa Uport 1110 Converter
  • Cigarette lighter receptacle power supply cables
  • AP2-1 Current Loop Unit
  • Barcode Scanners
  • Stands, wall mounting kits and mounting brackets
  • Displays

We use cookies to make your experience of our websites better.